Photometer PM 2111

Photometer

PM2111 model
The photometer “SOLAR” allows to research substrata, enzymes, electrolytes, hematologic indexes. It can be applied in laboratories of different branches. The photometer is applied in medicine as the semi-automatic biochemical analyzer in small and medium clinic diagnostic laboratories.

  • Possibility of choice and programming modes by the user
  • Possibility of use reagents at the choice of the user
  • Easy and convenient at using
  • Non-volatile storage
  • Built-in thermal printer

Cell section

Thermostatically controlled (37 °C) cell section, cell size: 12,5*12,5*45 mm, length of optical way is 10 mm

Programming of modes

The user can independently select, program and save in memory modes (up to 100 modes)

Open system

The photometer "SOLAR" works with reagents of different producers

Easy to use

Cozy LCD display: shows parameters of the selected mode, the end result of research, the hint during operation

Non-volatile storage

Non-volatile storage for 100 modes

Connection of the printer

Possibility of connection to the printer. Printing of research results
State registry SI of RB No. 03 25 0908 11
State registry SI of the Russian Federation No. 48521
State registry SI UKE of UA-MI/3p-216-2000

Measuring functions

  • Measurement of optical density and transmission
  • Two-wave measurements
  • Determination of concentration under factor and standard
  • Kinetic measurements

Delivery set

  • PM2111 photometer
  • Polystyrene one-use cells macro (1 ml) and/or semi-micro (0,4 ml)

By the additional order

  • Unit of samples preparation RT2110C
  • Polystyrene one-use cells macro (1 ml). Package is 100 pieces
  • Polystyrene one-use cell semi-micro (0,4 ml). Package is 100 pieces
  • Glass cell (1 ml)
  • Printer
Diagram of measurement single-channel
Light source halogen lamp
Wavelength 340, 405, 500, 540, 570, 620, 670 nanometers
Detector photo diode
Photometric range from -0,5 to 2,5 B; 0,4 to 100%Т
Fotometry error < 1% at D = 1,0 B
The volume of the researched sample 1,0 ml (macro); 0,4 ml (semi-micro)
Termostating temperature 37 °C, stability of kepping +-0,5 °C
Dimensions 335x235x125 mm
Weight 4,0 kg
Energy consumption 230 (+-10%) V, 50 Hz, 50 VA
Cell type Length of optical way , mm Spectral range, mm Internal sizes, mm Packing, pc
10031 One use polysterene MACRO cells 10 From 180 nm 10х10х44 100
10032 One use polysterene SEMI-MICRO cells 10 From 180 nm 10х4х44 100
10033 Glass cells 10 From 180 nm 10х10х43 1
Block of samples preparation PT 2110C
For preparation of samples by thermal stabilization
Nominal temperature of termostating 37 °C more
Sets of the light filters KCC-02 and KCC-04
For checking of the device
Overall dimensions of the light filter in frame of 12,5х12,5х45 mm more